EasyManua.ls Logo

Mindray uMEC10 - Page 82

Mindray uMEC10
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-10
Module Parameter
Low alarm limit High alarm limit
Auto alarm limits
range
Adult/
pediatric
Neonate
Adult/
pediatric
Neonate
NIBP-M
(Mean × 0.68
+ 8) mmHg
(Mean – 15)
or 35mmHg
(whichever
is greater)
(Mean × 0.86
+ 35) mmHg
(Mean + 15)
or 95 mmHg
(whichever
is smaller)
Adult: 30 to 230
Pediatric: 30 to 165
Neonate: 25 to 105
Temp
T1 (T1 0.5)ć
(T1 –
0.5) ć
(T1 + 0.5)ć (T1 + 0.5)ć 1 to 49 ć
T2 (T2 0.5)ć
(T2 –
0.5) ć
(T2 + 0.5)ć (T2 + 0.5)ć 1 to 49 ć
TD
Same as the
default alarm
limit
Same as the
default
alarm limit
Same as the
default alarm
limit
Same as the
default
alarm limit
Same as the
measurement range
IBP:
A
RT
/
A
o
U
AP/
BAP
FAP/
LV/
P1-P4
(Arterial
p
ressure)
IBP-S
(SYS × 0.68
ˇ10) mmHg
(SYS – 15)
or 45mmHg
(whichever
is greater)
(SYS × 0.86
ˇ38) mmHg
(SYS + 15)
or
105mmHg
(whichever
is smaller)
Adult: 45 to 270
Pediatric: 45 to 185
Neonate: 35 to 115
IBP-D
(Dia × 0.68 +
6)mmHg
(Dia – 15) or
20mmHg
(whichever
is greater)
(Dia × 0.86 +
32)mmHg
(Dia + 15) or
80mmHg
(whichever
is smaller)
Adult: 25 to 225
Pediatric: 25 to 150
Neonate: 20 to 90
IBP-M
(Mean × 0.68
+ 8)mmHg
(Mean – 15)
or 35mmHg
(whichever
is greater)
(Mean × 0.86
+ 35)mmHg
(Mean + 15)
or 95mmHg
(whichever
is smaller)
Adult: 30 to 245
Pediatric: 30 to 180
Neonate: 25 to 105
IBP: PA
IBP-S SYS × 0.75 SYS × 0.75 SYS × 1.25 SYS × 1.25
3 to 120mmHg IBP-D Dia × 0.75 Dia × 0.75 Dia × 1.25 Dia × 1.25
IBP-M Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25
IBP:
CVP/
ICP/
LAP/
RAP/
UVP/
P1-P4
(Venous
pressure)
IBP-M Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25 3 to 40mmHg