16-5
CAUTION
z To avoid blocking the airway, empty the DRYLINE II watertrap container
whenever half full. Replacing the DRYLINE II watertrap once a month is
recommended.
z The DRYLINE PRIME watertrap mini is automatically blocked when its service
life is reached. Replace the DRYLINE PRIME watertrap mini if the alarm “CO2:
Change Watertrap” occurs.
z The watertrap has a filter preventing bacterium, water and secretions from
entering the module. Extended use could destroy the filter in watertrap and fail to
stop the bacterium, water and secretions entering the module, result in damaging
the gas module and having infection risk..
NOTE
z Do not apply adult watertrap to the neonate patient. Otherwise, patient injury
could result.
z To extend the lifetime of the watertrap and module, disconnect the watertrap from
the module and set the operating mode to Standby mode when CO
2
monitoring is
not required.
z The sapmle rates are different when different types of watertraps are used.
16.5.2 Zeroing the CO
2
Module
The CO
2
module performs zero calibration automatically when needed.
NOTE
z The CO
2
module temporally stops measuring during zeroing.

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo
O2No

Related product manuals