EasyManuals Logo

Mindray uMEC10 User Manual

Mindray uMEC10
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #249 background imageLoading...
Page #249 background image
C-1
C Default Configurations
This chapter lists some of the most important factory default settings in configuration
management. You cannot change the factory default configuration itself. However, you can
make changes to the settings from the factory default configuration and then save the
changed configuration as a user configuration. The last column of the following tables is for
your notes and review.
Note: In this chapter, O.M means the monitor’s operating mode. Column C refers to the
settings that can be changed in configuration management. Column M refers to the settings
that can be changed in monitoring mode.
C.1 Parameters Configuration
C.1.1 ECG
ECG Setup
Item Name
O.M
General User Defaults
C M
Lead Set * * Auto
Alm Source * * HR
Alarm * * On
Alm Lev * * Med
HR/PR
High
Adu
* *
120
Ped 160
Neo 200
HR/PR
Low
Adu
* *
50
Ped 75
Neo 100
Sweep * * 25 mm/s
Beat Vol * * 2
Paced * No
Notch Filter * * On
Gain * * X1
Filter * * Monitor
ECG Display * * Normal

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo
O2No

Related product manuals