EasyManuals Logo

Mindray uMEC10 User Manual

Mindray uMEC10
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #75 background imageLoading...
Page #75 background image
7-3
7.3.2 Alarm Message
When an alarm occurs, an alarm message will appear in the technical or physiological alarm
area. For physiological alarms, the asterisk symbols (*) before the alarm message match the
alarm level as follows:
 High level alarms: ***
 Medium level alarms: **
 Low level alarms: *
Additionally, the alarm message uses different background color to match the alarm level:
 High level alarms: red
 Medium level alarms: yellow
 Low level physiological alarms: yellow
 Low level technical alarms: blue
You can view the alarm messages by selecting the physiological or technical alarm area.
7.3.3 Flashing Numeric
If an alarm triggered by an alarm limit violation occurs, the numeric of the measurement in
alarm will flash every second, and the corresponding alarm limit will also flash at the same
frequency indicating the high or low alarm limit is violated.
7.3.4 Audible Alarm Tones
The alarm tone is distinct from heart beat tone, keystroke tone and pulse tone in frequency.
This monitor has three choices of alarm tones and patterns: ISO, Mode 1 and Mode 2. For
each pattern, the alarm tones identify the alarm levels as follows:
 ISO pattern:
 High level alarms: triple+double+triple+double beep.
 Medium level alarms: triple beep.
 Low level alarms: single beep.
 Mode 1:
 High level alarms: high-pitched single beep.
 Medium level alarms: double beep.
 Low level alarms: low-pitched single beep.

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo
O2No

Related product manuals