20-5
20.6 Removing Paper Jam
If the recorder works incorrectly or produces unusual sounds, check if there is a paper jam
first. If a paper jam is detected, follow this procedure to remove it:
1. Open the recorder door.
2. Take out the paper and tear off the draped part.
3. Reload the paper and close the recorder door.
20.7 Cleaning the Recorder Printhead
If the recorder has been used for a long time, deposits of paper debris may collect on the
printhead compromising the print quality and shortening the lifetime of the roller. Follow this
procedure to clean the printhead:
1. Take measures against the static electricity such as Disposable Wrist Strap for the work.
2. Open the recorder door and take out the paper.
3. Gently wipe around the printhead using cotton swabs dampened with alcohol.
4. After the alcohol has completely been dried, reload the paper and close the recorder
door.
CAUTION
z Do not use anything that may destroy the thermal element.
z Do not add unnecessary force to the thermal head.

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo
O2No