8-2
NOTE
z After defibrillation, the screen display recovers within 10 seconds if the correct
electrodes are used and applied in accordance with the manufacturer’s instructions
for use.
8.3 Preparing to Monitor ECG
8.3.1 Preparing the Patient and Placing the Electrodes
1. Prepare the patient’s skin. Proper skin preparation is necessary for good signal quality at
the electrode, as the skin is a poor conductor of electricity. To properly prepare the skin,
choose flat areas and then follow this procedure:
 Shave hair from skin at chosen sites.
 Gently rub skin surface at sites to remove dead skin cells.
 Thoroughly cleanse the site with a mild soap and water solution. We do not
recommend using ether or pure alcohol, because this dries the skin and increases
the resistance.
 Dry the skin completely before applying the electrodes.
2. Attach the clips or snaps to the electrodes before placing them.
3. Place the electrodes on the patient.
4. Attach the electrode cable to the patient cable and then plug the patient cable into the
ECG connector.
8.3.2 Choosing AHA or IEC Lead Placement
1. Select the ECG parameter window or waveform area to enter the [ECG Setup] menu.
2. Select [Others][Lead Set] and then select [3-lead], [5-lead], [12-lead] (for
uMEC12/uMEC15) or [Auto] according to the applied electrodes.
3. Select [Main Menu][Maintenance >>][User Maintenance >>] enter the
required password
4. Select [Others >>][ECG Standard] and then select [AHA] or [IEC] according to
the standard that is applied for your hospital.

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo
O2No

Related product manuals