7-7
7.4.5 Setting the Reminder Tones
When the alarm volume is set to zero, or the alarm tone is reset or switched off, the patient
monitor issues a periodical reminder tone. To set the reminder tones:
1. Select [Main Menu][Maintenance >>][User Maintenance >>]enter the
required password.
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
3. Set the [Reminder Tones] to [On], [Off] or [Re-alarm]. When [Re-alarm] is selected,
the current physiological alarms and technical alarms marked with “Ĝ” will be
re-generated after the [Reminder Interval] if the alarm condition persists.
To set the interval between reminder tones, select [Reminder Interval] and toggle between
[1min], [2min] and [3min].
In addition, you can set the volume of alarm reminder tones. To set the volume of alarm
reminder tones, select [Main Menu][Alarm Setup >>][Others] or the [Alarm Setup]
QuickKey[Others]. Then, select [Reminder Vol] and toggle between [High], [Medium]
and [Low].
7.5 Understanding the Alarm Setup Menu
Select the [Alarm Setup] QuickKey or [Main Menu][Alarm Setup >>] to enter the
[Alarm Setup], where you can:
Set alarm properties for all parameters.
Change ST alarm settings.
Change arrhythmia alarm settings.
Set the threshold for some arrhythmia alarms.
Change other settings.

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo
O2No

Related product manuals