5-5
5.7 Transferring a Configuration
When installing several monitors with identical user configuration it is not necessary to set
each unit separately. An USB drive may be used to transfer the configuration from monitor to
monitor.
To export the current monitor’s configuration:
1. Connect the USB drive to the monitor’s USB port.
2. Select [Export Config. >>] in the [Manage Configuration] menu.
3. In the [Export Config.] menu, select the configurations and [User Maintenance
Settings] to export. Then select the [Export] button. A status message will report
completion of the transfer.
To import the configuration on the USB drive to the monitor:
1. Connect the USB drive to the monitor’s USB port.
2. Select [Import Config. >>] in the [Manage Configuration] menu.
3. In the [Import Config.] menu, select the configurations and [User Maintenance
Settings] to import. Then select the [Import] button. A status message will report
completion of the transfer.
5.8 Loading a Configuration
You may make changes to some settings during operation. However, these changes or the
pre-selected configuration may not be appropriate for the newly admitted patient. Therefore,
the monitor allows you to load a desired configuration so as to ensure that all the settings are
appropriate for your patient.
To load a configuration,
1. Select [Load Configuration >>] from the main menu.
2. The popup menu shows the existing configurations on the monitor. Selecting [Config.
on USB drive >>] will show the existing configurations on the USB drive.
3. Select a desired configuration.
4. Select [View] to view the configuration details. In the popup menu, you can select
[Alarm Setup >>], [Screen Setup >>] or [Parameter >>] to view the corresponding
contents. The alarm setup items which are different than those currently used are marked
in red.
5. Select [Load] to load this configuration.

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo