EasyManua.ls Logo

Mindray uMEC10 - Page 250

Mindray uMEC10
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C-2
Item Name
O.M
General User Defaults
C M
Smart Lead Off * * On
ST Analysis
Item Name
O.M General User Defaults
C M
ST Analysis * * Off
Alarm * * Off
Alm Lev * * Med
ST-X High * * when ST Unit is mV:
when ST Unit is mm:
0.20
2.0
ST-X Low * * when ST Unit is mV:
when ST Unit is mm:
-0.20
-2.0
ISO * * -80 ms
J 48 ms
ST J + 60 ms
X represents I, II, III, aVR, aVL, aVF, V, V1, V2, V3, V4, V5 or V6.
Arrh. Analysis
Item Name
O.M
General User Defaults
C M
Arrhythmia Threshold Settings
PVCs/min * *
Adu, Ped:
Neo:
10
10
Tachy * *
Adu:
Ped:
Neo:
120
160
180
Brady * *
Adu:
Ped:
Neo:
50
75
90
Asys. Delay * *
Adu, Ped:
Neo:
5
5
Vtac Rate * *
Adu, Ped:
Neo:
130
160

Table of Contents

Other manuals for Mindray uMEC10

Related product manuals