A.7.4 PR
Alarm limit Range (bpm) Step (bpm)
PR High (low limit +2) to 300
1
PR Low 15 to (high limit-2)
PR from SpO
2
Module
Measurement range 20 to 300 bpm
Resolution 1 bpm
Response time
< 30 s (PI > 0.3, no disturbance, PR value sudden change within
25 – 240 bpm)
Accuracy ±3 bpm
Refreshing rate 1 s
SpO
2
averaging time
7 s (when sensitivity is set to High)
9 s (when sensitivity is set to Medium)
11 s (when sensitivity is set to Low)
PR from IBP Module
Measurement range 25 to 350 bpm
Resolution 1 bpm
Accuracy ±1 bpm or ±1%, whichever is greater
Refreshing rate 1 s
A.7.5 NIBP
Standards Meet standards of IEC80601-2-30,
Technique Oscillometry
Mode of operation Manual, Auto and STAT
Auto mode repetition intervals
1 min, 2 min, 2.5 min, 3 min, 5 min, 10 min, 15 min, 20 min, 30
min, 1 h, 1.5 h, 2 h, 3 h, 4 h, 8 h
STAT mode cycle time 5 min
Max measurement time
Adult, pediatric: 120 s
Neonate: 90 s
Measurement ranges
(mmHg)
Adult Pediatric Neonate
Systolic: 25 to 290 25 to 240 25 to 140

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo
O2No

Related product manuals