A-13
*Measurement accuracy verification: In adult and pediatric modes, the blood pressure
measurements measured with this device are in compliance with the Standard for
Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and stardard
deviation by comparing with intra-arterial or auscultatory measurements (depending on the
configuration) in a typical patient population. For auscultatory reference, the 5th Korotkoff
sound was used to determine the diastolic pressure.
In neonatal mode, the blood pressure measurements measured with this device are in
compliance with the American National Standard for Non-invasive sphygmomanometers
(ISO 81060-2) in terms of mean error and stardard deviation by comparing with
intra-arterial measurements (depending on the configuration) in a typical patient population.
A.7.6 Temp
Standards Meet standard of ISO 80601-2-56
Technique Thermal resistance
Operating mode Direct mode
Measurement range 0 to 50 ć (32 to 122 )
Resolution 0.1 ć
Accuracy ±0.1 ć(without probe)
Refreshing rate 1 s
Minimum time for accurate
measurement
Body surface: <100 s
Body cavity: <80 s
Transient response time
Body surface probe : <45 s
Body cavity probe: <45 s
Alarm limit Range Step
T1/T2 High (low limit +1) to 50 ć
0.1 ć
T1/T2 Low 0.1 to (high limit -1) ć
TD High 0.1 to 50 ć