C-7
C.1.6 NIBP
Item Name
O.M
General User Defaults
C M
Alarm * * On
Alm Lev * * Med
Interval * * 15 min
NIBP End Tone * * Off
Cuff Press.
˄mmHg˅
Adu
* *
80
Ped 60
Neo 40
Initial
Pressure
˄mmHg˅
Adu
* *
160
Ped 140
Neo 90
Alarm Limits
NIBP-S High
˄mmHg˅
Adu
* *
160
Ped 120
Neo 90
NIBP-S Low
˄mmHg˅
Adu
* *
90
Ped 70
Neo 40
NIBP-M High
˄mmHg˅
Adu
* *
110
Ped 90
Neo 70
NIBP-M Low
˄mmHg˅
Adu
* *
60
Ped 50
Neo 25
NIBP-D High
˄mmHg˅
Adu
* *
90
Ped 70
Neo 60
NIBP-D Low
˄mmHg˅
Adu
* *
50
Ped 40
Neo 20

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo
O2No