D-2
Measurement Alarm messages L Cause and solution
Extreme Brady H
R on T M*
Run PVCs L*
PVCs/min M*
Multif.PVC M*
Bigeminy M*
Trigeminy M*
Tachy M*
Brady M*
Vent. Rhythm M*
Multif. PVC M*
Nonsus. Vtac M*
Pause L*
Resp
Resp Apnea H
The respiration signal was so weak that the
monitor cannot perform respiration analysis.
Check the patient’s condition and the Resp
connections.
Resp Artifact H
The patient’s heartbeat has interfered with
his respiration. Check the patient’s condition
and the Resp connections.
SpO
2
SpO
2
Desat H
The SpO2 value has fallen below the
desaturation alarm limit. Check the patient’s
condition and check if the alarm limit
settings are correct.
No Pulse H
The pulse signal was so weak that the
monitor cannot perform pulse analysis.
Check the patient’s condition, SpO
2
sensor
and measurement site.
CO
2
CO
2
Apnea H
The patient stops breathing, or the
respiration signal was so weak that the
monitor cannot perform respiration analysis.
Check the patient’s condition and the RM
connections.

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo
O2No

Related product manuals