D-5
Measurement Alarm message L I Cause and solution
NIBP NIBP Loose Cuff L A The NIBP cuff is not properly connected, or
there is a leak in the airway.
NIBP Air Leak L A
NIBP Pneumatic Leak L A Check the NIBP cuff and pump for
leakages.
NIBP Cuff Type
Wrong
L A The cuff type applied mismatches the
patient category. Verify the patient category
and replace the cuff.
NIBP Air Pressure Err L A An error occurred to the air pressure. Verify
that the monitor application site meets the
environmental requirements and check if
there is any source that affects the air
pressure.
NIBP Weak Signal L A The patient’s pulse is weak or the cuff is
loose. Check the patient’s condition and
change the cuff application site. If the error
persists, replace the cuff.
NIBP Signal Saturated L A The NIBP signal is saturated due to excess
motion or other sources.
NIBP Overrange L A The measured NIBP value exceeds the
module measurement range.
NIBP Excessive
Motion
L A Check the patient’s condition and reduce the
patient motion.
NIBP Cuff Overpress. L A The NIBP airway may be occluded. Check
the airway and measure again.
NIBP Equip Err H A An error occurred during NIBP
measurement and therefore the monitor
cannot perform analysis correctly. Check
the patient’s condition and NIBP
connections, or replace the cuff.
NIBP Timeout L A
NIBP Measure Failed L A
NIBP Illegally Reset L A An illegal reset occurred during NIBP
measurement. Check if the airway is
occluded.
IBP YY Sensor Off L* A Check the sensor connection and reconnect
the sensor.
YY Disconnected H C The liquid way is disconneted from the
patient, or the three-way valve is open to the
air. Check the connection of the liquid way,
or check the valve is open to the patient. If
the problem remains, contact the Customer
Services Dept. for help.

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo
O2No

Related product manuals