D-8
Measurement Alarm message L I Cause and solution
Recorder Vlt Low L C
supply. Restart the monitor.
Recorder Head Hot L C The recorder has been working for too long
time. Stop the recording and resume the
recording till the recorder’s printhead cools
down.
Rec Paper Wrong Pos. L A Re-load the recorder paper.
System System Watchdog Err H C An error occurred to the system. Restart the
monitor.
System Software Err H C
System CMOS Full H C
System CMOS Err H C
System FPGA Err H C
System Err N H C
N is within 2 to 12.
Other Bed
Disconnected
L A Check network connection.
PWR interrupted.
Check meas. state
L A Power supply failed accidently. Check the
measurements when the monitor restarts.
No CMS L A The monitor is disconnected from the CMS.
Check network connection.

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo
O2No