32
3 Advanced Operations and Useful Functions
No. 99MAM029A
1
Set the WORK MODE to "MODE3".
1
While pressing
DATA
press
POWER
,
then just release
POWER
, when the print-
ing starts, then release
DATA
.
» The power is turned on and enters the
parameter setup mode.
2
Press
DATA
twice.
» The WORK MODE currently set is
printed.
2
3
3
Press
STAT.
OUT LOG
repeatedly until "3 WORK
MODE :MODE3" is printed.
4
Press
DATA
.
» "MODE3" is determined.
5
Press
DATA
repeatedly until the param-
eter settings list is printed.
» Parameter setup mode is completed.
This completes the parameter setting.
Next, move to the subgroup measurement.
When the WORK MODE is set to "MODE3", the key operation differs from usual as
shown below.
Key
During subgroup
measurement
After completion of
subgroup measurement
CLEAR
Re-input from the No.1 data. Delete all the measurement data (setting con-
tents will remain).
CANCEL
Cancel the previously inputted
measurement data.
Deletes the subgroup for which input was previ-
ously completed.
TOL.
REC/STOP
Stop measurement and release
the measuring mode.
Move to the next subgroup measurement.
STAT.
OUT LOG
Calculates X¯ and R to print the
result completing the subgroup
measurement.
Calculates the control limits with all the sub-
groups' data input up to that point of time and
prints the results.
POWER
TOL.
REC/STOP
PRINTER
CLEAR
STAT.
OUT LOG
CANCEL
DATA
FEED