18
3 Parameter Setting
No. 99MBC122A
3.2
List of Parameters
No. Setting item
Description/Set values
(The values in bold indicate initial values.)
00 Parameter mode Used for setting/saving/loading parameters or setting the optional constant
value.
0: Setting parameters
1: Setting CEL-specic parameters
2: Setting optional constants
3: Saving parameters
4: Loading parameters
Tips
For details about saving and loading parameters, see "5.3.3 Saving and Loading
Parameters" (page 44).
01 Parameter initialization If you set the value of this parameter to 1, the set values for all parameters,
except for the unit setting, can be reset to their factory defaults.
Once this setting has been enabled, this parameter is initialized to reset the
set value to 0 (disable).
0: Disabled (one shot)
1: Clear (reset to the initial values)
02 Key protect Disables key operations to prevent operation errors.
0: Normal
1: Disable key inputs
05 (Modication prohib-
ited)
0:-
1: (Not available)
06 Display mode selection Selects the UNIT (counting method) that is assigned to each CEL. The UNIT
to be set can be selected as desired.
0: Two-coordinate display
1: Calculation display
2: Dual display
3: Memory display
4: Speed display
5: Optional one-axis display
6: Optional two-axis display
7: Optional four-axis display
Tips
For details about the display modes, see "3.3.2 Details of Display Mode
Selection (Parameter Number 06)" (page 24).
07 Display at startup Selects whether to display the standby state or the counter display at startup.
0: [- - - -] display
1: 0.000