LS-50
・Density expressible range:
The lm with 2.5 - 3.0 density should be scanned within the gradient range.
・Prescan reproducibility
The color variation of 24-patch Lab should be ΔE4 or less. (Reference value: average of 5-times
scanning)
・Drift
The average ΔE of 24-patches should be 5 or less; ΔE for every patch should be 10 or less.
The shading standard should be met.
・Streaking
No streaks in both main- and sub-scanning directions.
・Grain roughness
Indistinctive. No signicant difference from the traditional models.
・Irregular color
Color difference (ΔE) should be 5 or less.
・Inuence of outside light
No inuence
7. Operating sounds
No signicant abnormal sounds with 55-dB or less at a 1-m distance away.
8. Status indication
BUSY: 0.635Hz ± 1% of blinking frequency
ERROR: 5Hz ± 1% of blinking frequency
9. Force to attach/remove
・Attach adapter:
MA-21 500 ~ 1500 gf
IA-20(S) 500 ~ 5000 gf
SA-21 500 ~ 6000 gf
・Remove adapter:
MA-21 500 ~ 1800 gf
IA-20(S) 500 ~ 5000 gf
SA-21 500 ~ 5000 gf