VRA53801-R.3609.A
- A29 ・ -
LS-50
How to use Inspection Software
1. Purpose
This software is used when LS-5000ED is disassembled or its parts are replaced or inspections are made.
2. Preparation (tools required)
・DOS/V PC
・Inspection chart for (J61187)
・Positive lm chart(J61189); Insert this into MA-21
・IR pass lter (J63088); Insert this into MA-21
・Inspection software (J65050A: Japanese, J65050B: English)
【Hardware requirements】
・Windows2000
・Japanese/English OS
・CPU Pentium Ⅲ 700MHz or more
・RAM 256MB or more
3. Inspection details
1) Initialization
・Check the rmware version.
2) Illumination irregularity
・Check for the raw data of the CCD with shading OFF in the standard range.
3) Shading
・Check for the read-in data of the CCD waveform with shading ON in the standard range.
4) Noise inspection
・By performing prescanning operation, check if the operating sound has no problem.
5) Mechanical precision
・ Scanning position accuracy
Get the central position (average values of x, y) of each alignment mark at 4 positions, then comparing them with set
values, check the deviation of image scanning position (shading).