EasyManua.ls Logo

Optek CONTROL 200 - Page 86

Optek CONTROL 200
135 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C200 SOFTWARE
Instruction Manual C200 Part2 A1.doc
optek Danulat GmbH D-45356 Essen Germany
page 64
Temperature Value – Point 1
Deg. C (or Deg. F) and Adjust are shown alternately on the upper display line.
Use the and keys to adjust the displayed reading to the required temperature
value for point 1 on the curve.
Ratio Value – Point 1
Ratio and Adjust are shown alternately on the center display line. Use the and
keys to adjust the displayed reading to the required ratio value for point 1 on the
curve.
Temperature and Ratio Values – Points 2 to 6
Enter the temperature and ratio values for the remaining points in the user-
defined temperature compensation curve.
Filter Time
Set the required filter time, between 1 and 60 seconds in 1 second increments.
Sensor B configuration (dual input analyzers only) is identical to
Sensor A configuration.
See Section 5.2.4.
0.0
1.79
A: Enter point 1
Deg.C
Ratio
1.79
A: Enter point 1
Deg.C
User Temp Comp
set to
Yes
0.0
Ratio
125.0
0.30
A: Enter point 6
Deg.C
Ratio
A: Filter Time
10
Secs
Config. Sensor A
CONFIG. ALARMS
Config. Sensor B

Table of Contents

Related product manuals