8.3.5.1 System commands
System commands are used for system-level settings.
COMMAND OPTIONS Read/Write DESCRIPTION
SYS CAT R
Returns a list of existing
hardware devices
SYS DISP DIMA OFF:ON R/W Set automatic display brightness
SYS DISP DIMT 0 - 10 R/W
Sets the time for the automatic
dimming
SYS DISP BRIG 0 - 100 R/W Sets the brightness
SYS MAN HVER R
Reads Hardware version of the
Box
SYS MAN FVER R Cryosys version
SYS MAN SERL R Serial number of the box
SYS TIME hh:mm:ss R/W Sets the time of the box
SYS DATE yyyy:MM:dd R/W Sets the date of the box
SYS USER NONE:GUEST:NORM:ENG R/W
Sets the remote user level (a
password may be required after
the level)
SYS PASS W
Sets a new password for the
present user level
SYS LOCK OFF:SOFT:ON W
Sets the lock mode on the unit
when in remote mode
SYS RST W Resets the hardware
SYS FLSH R
Reads the available memory in
the box
SYS RAM R
Reads the available RAM in the
box
SYS ALRM R Lists current alarms
SYS RUN Command Name W
Runs pre-loaded system
commands
SYS GUI CHAN UID signal name R/W
Set the signal display for a given
channel widget
SYS GPIB ADDR 1- 31 R/W Sets the GPIB address of the box
SYS TCP ADDR #.#.#.# R/W Sets the IP address of the box
SYS TCP DHCP OFF:ON R/W Sets DHCP search
SYS TCP GATE #.#.#.# R/W Sets the gateway
SYS TCP SUBN #.#.#.# R/W Sets the subnet mask
SYS RS232 ADDR 1-255 R/W Sets the Isobus address
SYS RS232 BAUD 9600:19200:38400:57600:115200 R/W Sets the Baudrate
SYS RS232 STOP 1 - 2 R/W Sets the number of stop bits
SYS RS232 DATA 7 - 8 R/W Sets the number of data bits
SYS RS232 PAR none:odd:even:mark:space R/W Sets the parity
SYS RS232 FLOW none:hardware:Xon/Xoff R/W Sets the flow control