Page 106
A – Amps, V – Volts or W – Watts.
8.3.5.1 System commands
System commands are used for system-level settings.
COMMAND OPTIONS Read/Write DESCRIPTION
SYS CAT R
Returns a list of existing
hardware devices
SYS DISP DIMA OFF:ON R/W Set automatic display brightness
SYS DISP DIMT 0 - 10 R/W
Sets the time for the automatic
dimming
SYS DISP BRIG 0 - 100 R/W Sets the brightness
SYS MAN HVER R
Reads Hardware version of the
Box
SYS MAN FVER R Cryosys version
SYS MAN SERL R Serial number of the box
SYS TIME hh:mm:ss R/W Sets the time of the box
SYS DATE yyyy:MM:dd R/W Sets the date of the box
SYS USER NONE:GUEST:NORM:ENG R/W
Sets the remote user level (a
password may be required after
the level)
SYS PASS W
Sets a new password for the
present user level
SYS LOCK OFF:SOFT:ON W
Sets the lock mode on the unit
when in remote mode
SYS RST W Resets the hardware
SYS FLSH R
Reads the available memory in
the box
SYS RAM R
Reads the available RAM in the
box
SYS ALRM R Lists current alarms
SYS RUN Command Name W
Runs pre-loaded system
commands
SYS GUI CHAN UID signal name R/W
Set the signal display for a given
channel widget
SYS GPIB ADDR 1- 31 R/W Sets the GPIB address of the box
SYS TCP ADDR #.#.#.# R/W Sets the IP address of the box
SYS TCP DHCP OFF:ON R/W Sets DHCP search
SYS TCP GATE #.#.#.# R/W Sets the gateway
SYS TCP SUBN #.#.#.# R/W Sets the subnet mask
SYS RS232 ADDR 1-255 R/W Sets the Isobus address
SYS RS232 BAUD 9600:19200:38400:57600:115200 R/W Sets the Baudrate
SYS RS232 STOP 1 - 2 R/W Sets the number of stop bits
SYS RS232 DATA 7 - 8 R/W Sets the number of data bits
SYS RS232 PAR none:odd:even:mark:space R/W Sets the parity
SYS RS232 FLOW none:hardware:Xon/Xoff R/W Sets the flow control

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Oxford Instruments Omicron NanoScience MercuryiTC and is the answer not in the manual?

Oxford Instruments Omicron NanoScience MercuryiTC Specifications

General IconGeneral
BrandOxford Instruments Omicron NanoScience
ModelMercuryiTC
CategoryController
LanguageEnglish