9 Run some standard procedures to test that the iTC is working as expected.
10 With the iTC in Local mode and if the new firmware is acceptable:
a) Tap Settings on the Home page.
b) Select the Updates tab.
c) Tap Save Changes to accept the new firmware version.
11 With the iTC in Local mode and if the new firmware is not acceptable:
a) Tap Settings on the Home page.
b) Select the Updates tab.
c) Tap Revert Changes to revert to the old firmware version.
12 Power the system down and wait 30 seconds before applying power again.
6.7 Access Level
This page allows the user to enter ENGINEERING mode, to access features that are not
available to normal users. The default engineering password is available on request but should
be used with caution. In ENGINEERING mode it is possible to change all the system
parameters and damage could occur on some systems if these parameters are changed to
inappropriate values. The user can change this password, if desired (see below).
Entering engineering mode
1 Tap Settings on the Home page.
2 Scroll right until the Access Level tab is visible.
3 Tap Access Level. The engineering mode page opens.