Operator's Manual
Issue 05 /
Feb 2015 / UMC0071
MercuryiTC
©2014 Oxford Instruments Omicron NanoScience. All rights reserved.
Page 10
1.1.2 High voltage hazard
HIGH VOLTAGE
Danger of death or serious injury. Switch off the external AC
electrical supply and disconnect and remove the external supply
cable, before maintaining this equipment or removing any covers.
HAUTE TENSION
Danger de mort ou de blessure grave. Sectionner l'alimentation
électrique C.A. externe puis déconnecter et retirer le cordon
d'alimentation externe avant d'effectuer toute maintenance de cet
équipement ou d'en retirer les capots.
This equipment contains high voltages. Isolate this equipment before removing and covers. To
isolate this equipment, switch off the external AC electrical supply, and disconnect and remove
the external supply cable.
The AC electrical supply is considered to be the disconnect device for the equipment. Do not
restrict access to the AC electrical supply at any time. The AC electrical supply cable must
remain accessible for disconnection of the equipment.
1.1.3 Maintenance and adjustment
Only qualified and authorised persons must service or repair this equipment.
Under no circumstances must the user adjust or repair this equipment while the electrical power
supply is connected.
Ensure that the equipment is disconnected from the AC electrical power supply before the
covers are removed or fuses are removed or fitted. It is NOT sufficient to turn off the switch on
the rear panel or the < On/Off > button on the front panel.
The equipment contains capacitors that may remain charged after AC electrical power is
removed. Discharge all capacitors before starting work.
1.1.4 Restrictions on use
The equipment is not suitable for use in explosive, flammable or hazardous environments.
The equipment does not provide protection against the ingress of water. The equipment must
be positioned so that it will not be exposed to water ingress.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Oxford Instruments Omicron NanoScience MercuryiTC and is the answer not in the manual?

Oxford Instruments Omicron NanoScience MercuryiTC Specifications

General IconGeneral
BrandOxford Instruments Omicron NanoScience
ModelMercuryiTC
CategoryController
LanguageEnglish