EasyManua.ls Logo

Oxford Instruments Omicron NanoScience MercuryiTC - Resistance Measurement in Constant Voltage Mode

Oxford Instruments Omicron NanoScience MercuryiTC
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator's Manual
MercuryiTC
©2014 Oxford Instruments
Omicron NanoScience
Rsensor = Rref x | ADCnorm
| x
Where:
Rref
is the reference resistance (either 2 kohms or 80 kohms).
ADCnorm
is the ADC reading, normalised to the range
RefGain
is the interpolated gain from the calibration tables.
ADCgain
is the gain setting of the ADC.
The measurement accuracy of this circuit does not depend on the accuracy of the excitation
current or the reference voltage. The accuracy prim
reference resistor plus any errors introduced by operational amplifiers.
9.3.3 Resistance
measurement
Constant voltage measurement mode is used with sensors that have a negative temperature
coefficient.
The sensor and a reference resistor (either R44 or R45) are connected in series. Digital to
analogue converters U24 and U25 generate an upper and lower demand voltage. The upper
voltage is buffered by U10 and applied to the upper end of the sensor resistor
Issue 05 /
Feb 2015 / UMC0071
Omicron NanoScience
. All rights reserved.
| x
RefGain / ADCgain
is the reference resistance (either 2 kohms or 80 kohms).
is the ADC reading, normalised to the range
0 to ±1 V.
is the interpolated gain from the calibration tables.
is the gain setting of the ADC.
The measurement accuracy of this circuit does not depend on the accuracy of the excitation
current or the reference voltage. The accuracy prim
arily depends on the accuracy of the
reference resistor plus any errors introduced by operational amplifiers.
measurement
in
constant voltage mode
Constant voltage measurement mode is used with sensors that have a negative temperature
The sensor and a reference resistor (either R44 or R45) are connected in series. Digital to
analogue converters U24 and U25 generate an upper and lower demand voltage. The upper
voltage is buffered by U10 and applied to the upper end of the sensor resistor
Feb 2015 / UMC0071
Page 123
The measurement accuracy of this circuit does not depend on the accuracy of the excitation
arily depends on the accuracy of the
constant voltage mode
Constant voltage measurement mode is used with sensors that have a negative temperature
The sensor and a reference resistor (either R44 or R45) are connected in series. Digital to
analogue converters U24 and U25 generate an upper and lower demand voltage. The upper
voltage is buffered by U10 and applied to the upper end of the sensor resistor
. The lower

Table of Contents