TARGAS-1 Operation Manual V. 1.02 102 support@ppsystems.com
2
maximum. This is the same value as the Max CO
2
value for the Graph setting and can be modified under Main >
Settings > Settings 2 > Graph.
Stops the Injection process and returns to the Injection Baseline
menu (Step 4). Process results are not saved to the USB flash drive.
Stops the Injection process and returns to Injection Baseline menu
(Step 4). Injection Process results are saved to the USB flash drive.
Typically during an injection, the first value will rise quickly and drop back to zero; correspondingly, the
value of M will increase quickly as the value of C peaks and then plateaus at a constant value. Once C
has dropped to 0 ppm (or a very low background value), wait 3-5 seconds to make sure M has stabilized
and take note of the value. This value is also recorded on the USB memory stick, if a memory stick has
been inserted.
Errors from injection measurements are typically larger than flow-through methods (larger errors could
come from both instrumental and handling sources), so it is recommended that you measure the same
sample multiple times and calculate an average whenever possible.
Information Menu
Provides information about serial numbers and the firmware versions
Information to contact PP Systems
Returns to the Main Menu.