TARGAS-1 Operation Manual V. 1.02 81 support@ppsystems.com
CPY ā Start Process Menu
Returns to the Processes Menu.
Confirms and starts the CPY process by displaying the CPY ā Volume and
Area Settings menu (Step 2).
CPY ā Volume and Area Settings (Step 2)
The volume and area settings are defined in this screen.
Important Note. The above values are based on the default volume for the CPY-5 Canopy Assimilation
Chamber only. You must remember to account for any volume changes in order for the instrument to
properly calculate assimilation rates. If using collars, please make sure that you adjust the volume
accordingly.
CPY ā Volume and Area Settings Menu
Sets the volume and area values back to the defaults.
Sets the volume of the system. Entries between 1,000 and 5,000 ml are
permitted. Default is 2427 ml.
Sets the surface area of the chamber in cm
2
. Entries between 150 and 200
cm
2
are permitted. The default value is 167 cm
2
This value is automatically updated when there is a change to either the
Volume or Area values.
Stops the CPY process and returns to the Processes Menu.