TARGAS-1 Operation Manual V. 1.02  34  support@ppsystems.com 
 
•  Type: Molecular Sieve, 13X 1/16, 1.25 lb. container 
•  Manufacturer: AGM Container. (www.agmcontainer.com
) 
•  For the latest Material Safety Data Sheet, please visit www.agmcontainer.com and 
request the latest MSDS or contact PP Systems. 
Take caution to wash your hands completely after handling Molecular Sieve. 
Molecular Sieve, 1.25 lb. 
 
Molecular Sieve Repackaging 
The Molecular Sieve is originally supplied by PP Systems in tin packaging.  After 
initial opening, we strongly urge all users to repackage the Molecular Sieve in 
small glass containers with a screw top to seal the desiccant from room air. This 
desiccant saturates very quickly in room air and if not properly stored it will cause 
it to go bad and subsequently affect CO
2
 readings and calibration. 
To ensure a good seal, we also recommend putting some electrical tape around 
the screw top as shown here.  If you have any questions, get in contact with PP 
Systems. 
 
 
 
CO
2
 & H
2
O Control Columns 
There are two columns used for CO
2
 and H
2
O control.  One column contains soda lime (CO
2
 scrubber) 
and the other contains Drierite (H
2
O scrubber). 
Soda Lime 
Soda lime (calcium hydroxide, sodium hydroxide, water) is used to remove CO
2
 from air entering the 
TARGAS-1. Both self-indicting (white to violet) and non-indicating Soda Lime can be can be used with the 
TARGAS-1.  Soda Lime cannot be regenerated and should be discarded after exhaustion. 
•  Type: Sofnolime, 1.0-2.5 mm, self-indicating white to violet), 1 kg 
•  Manufacturer: Molecular Products. (www.molecularproducts.com
) 
Tip 
Always change the molecular sieve at least 
once per week when the TARGAS-1 is in use 
regardless of operation time.