TARGAS-1 Operation Manual V. 1.02 23 support@ppsystems.com
PLC5 Leaf Cuvette
The PLC5 Leaf Cuvette is designed to work with the TARGAS-1 Portable CO
2
/H
2
O Analyzer for
measurement of leaf gas exchange on a wide variety of plants including broad leaves, narrow leaves,
grasses and small needle conifers.
When the PLC5 Leaf Cuvette is not in use or being stored away make sure that the Open/Close Head
Latch is secure in place to keep the cuvette head open to avoid problems associated with compressed
leaf gaskets which is a common source for leaks.
Light Unit (Optional)
The optional light unit can be used with the PLC5
Leaf Cuvette for use on cloudy days or for light
response curves. It easily clips onto the PLC5 Leaf
Cuvette head and provides control of light intensity
up to 2500 µmol m
-2
s
-1
.
The electrical connection is
on the side of cuvette handle as shown above.
• Type: LED (white)
• Measurement Range: 0-2500 µmol m
-2
s
-1
Head
2
Release Screw
Electrical
Switch