TARGAS-1 Operation Manual V. 1.02 138 support@ppsystems.com
Error Messages
Error and status messages are displayed in the Status Area during system operation. The following table
highlights the most common messages that you may see. An error message may appear after making a
change to system setting and if this happens do not be concerned as the message will likely disappear
after a few seconds. However, if an error message is displayed and does not disappear after a few
seconds then the instrument is telling you something and the message should be addressed.
A/D values from the last Zero are too low (<2500).
2
2
setting (250 ppm by
default).
2
is greater than the high CO
2
setting (2000 ppm by
default).
Analyzer temperature < 50
o
C.
Analyzer temperature > 60
o
C.
Sample flow rate is less than 50 cc/min.
IRGA unable to perform a write operation. If error persists, call PP
Systems technical support.
IRGA read/write EEPROM CRC incorrect. If error persists, call
PP Systems technical support.
IRGA setting error. Value is out of range. If error persists, call PP
Systems technical support.
Reached Process CO
2
limit
Reached Process time limit
The SR measurement is Non Linear
The process has been terminated
The Absorber column material has been exhausted and needs
replacing
Zero will occur in 2 minutes
Zero will occur in 1 minutes
Could not successfully write the IRGA DigiPot
Recording a record to the memory stick
Host only. First character is not recognized as a command.
Host only. First character is a recognized command but the next
character is invalid. Example: S1,1<CR> (as opposed to
Host only. Command is correct but parameter is not valid.
Example: S,20,5. The value of 20 is an invalid parameter.
Host Only. Value entered is below the valid range of values for
that parameter.
Host Only. Value entered is above the valid range of values for
that parameter.
Plc-3 internal par 1 error
Plc-3 internal par 2 error
Expecting a PLC3 to be connected to port 2
Expecting a TARGAS PLC5 to be connected to port 2
Expecting a PAR2 device to be connected to port 1
Expecting a STP-2 device to be connected to port 1
Expecting a SRC device to be connected to Port 1 or Port 2