TARGAS-1 Operation Manual V. 1.02  108  support@ppsystems.com 
 
 
 
2
O concentration (mb) of the calibration reference gas. 
Entering a value will begin the Calibration Process 
The current H
2
O reading based on the Scale Factor 
The current Scale Factor being used 
Returns to the Calibration Menu. 
 
Connecting Dew Point Generator to the TARGAS-1 
During calibration, the H
2
O IRGA readings are referenced against air of known H
2
O concentration.  For 
H
2
O recalibration we recommend using an accurate dew point generator or water vapor generator which 
are commercially available.  For more information please contact PP Systems. 
The TARGAS-1 calibration procedure uses two reference points: 
1.  Zero H
2
O, generated from air that has been scrubbed of H
2
O through the Molecular Sieve in the 
absorber column. 
2.  Span (reference) gas, a user-provided air with a known H
2
O concentration, typically from a very 
accurate (+/- 1%) source. 
 
Follow these instructions to calibrate the TARGAS-1 H
2
O sensor: 
1.  Connect the reference (sample) gas from your H
2
O calibration device to the GAS IN  port on the 
TARGAS-1. A T-fitting should be in place between your calibration device and the TARGAS-1 
console and left open in order to relieve excess pressure. 
 
2.  Set the flow on your device so that the flow rate exhausting out the T-fitting should be approximately 
250 ml/min. 
 
3.  Select the button to the right of the “H
2
O
 
Conc (mb):” text.  
 
4.  Enter the H
2
O concentration of the reference gas. This value must match the H
2
O concentration of 
the reference source and be greater than or equal to the anticipated measurement range of the 
instrument. For example, if the anticipated measurement range of your samples will be 0 - 20 mb, the 
instrument should be calibrated using reference air with an H
2
O concentration of 20 mb or higher.