TARGAS-1 Operation Manual V. 1.02 109 support@ppsystems.com
5. The TARGAS-1 will then perform a zero. After completion of successful zero, a similar message to
the following will be displayed.
Note, the displayed value may not be the value you entered in step 4 above. When the Actual H
2
O
value has stabilized (i.e., not changing by more than 0.1 mb after seconds), press the OK button. To
cancel without performing a Calibration, press Stop.
Calibration
Set H2O
The H
2
O concentration Value entered at the beginning of the Calibration Process.
Actual H2O
The current H
2
O Reading
Stop Button
Returns to the Calibration Menu with no changes.
OK Button
Calculates new Scale Factor and goes to next screen
If the process is successful, the screen below will be displayed and the new calibration constant will be
stored in non-volatile memory. If the process is unsuccessful, the message “Calibration: failed.” will be
displayed and no changes made.