PowerMeasuringModule PMM
35,0(
6
88
Revision 13 EN - 02/2023
17 Technical data
Measurement parameters Standard absorber
1)
Advanced absorber
1)
Power range 400 – 6 000W
2)
400–12000W
2)
Wavelength range BG 440–460nm / 510–540nm
NIR 800–1 100nm
Max. beam diameter 30mm
Max. power density on
the absorber (approx. 25
mm below the top edge
of the device when the
shutter is open) at beam
diameters
> 10mm 1.5kW/cm² 4kW/cm²
10 – 3mm 2.5kW/cm² 5kW/cm²
3 – 1.5mm 5kW/cm² 10kW/cm²
< 1.5mm 6kW/cm² 12kW/cm²
Irradiation time ²
)
0.1–2.0s (depending on laser power)
0.1 – 1.0 s (Type AP3s)
Min.
ontime/offtime (duty cycle)
for pulsed lasers
50 µs
(e.g. max. 10kHz at 50% duty cycle)
Max. laser rise time <1% of irradiation time
Energy per measurement 400 – 3500J
Recommended energy per measure-
ment
400 – 1000J
Total duration until measurement value
output
<15s
Total duration until measured value
output for devices with reduced
thermalization time (type AP3s)
<5s
Nominal measurement frequency 500J: 1cycle/ 2min
1000J: 1cycle/ 4min
1)
Please read the information on the identification plate to determine if your device is equipped with
a standard or advanced absorber.
2)
The stated limit values are to be understood in correlation with the permitted maximum energy (E= P∙t).
Device parameters Standard absorber
1)
Advanced absorber
1)
Max. absorber temperature 120°C
Max. angle of incidence perpendicular to
inlet aperture
±10°
Max. center tolerance ±2.0mm
Measuring accuracy at angles of inci-
dence up to 10°
±3%
Reproducibility ±1%