EasyManuals Logo

RKC INSTRUMENT RB100 User Manual

RKC INSTRUMENT RB100
306 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #275 background imageLoading...
Page #275 background image
10. SPECIFICATIONS
IMR02C15-E4
10-3
Sampling cycle: 250 ms
Influence of external resistance:
Approx. 0.25 µV/Ω (Converted depending on TC types)
Influence of input lead: Approx. 0.02 %/Ω of span (Only RTD)
10 Ω or less per wire
Input impedance: TC input: 1 MΩ or more
Voltage input: Approx. 1 MΩ
Current input: Approx. 250 Ω
(Connect a 250 Ω shunt resister to the input terminals)
Sensor current: Approx. 200 µA (Only RTD)
Action at input break: TC input: Upscale or downscale
(Select one of these)
RTD input: Upscale
Voltage input: Downscale or Indicates the value near 0
Current input: Downscale or Indicates the value near 0
Action at input short circuit:
Downscale (RTD input)
Input correction: PV bias: −1999 to +9999 °C [°F] or −199.9 to +999.9 °C [°F]
(TC/RTD)
−Input span to +Input span (Voltage/Current input)
PV digital filter (First order lag digital filter):
0 to 100 seconds (0: Filter OFF)
îš„ Current transformer (CT) input [optional]
Number of input: 2 points
CT type: CTL-6-P-N or CTL-12-S56-10-N (Sold separately)
Input range: CTL-6-P-N: 0.0 to 30.0 A
CTL-12-S56-10L-N: 0.0 to 100.0 A
Sampling cycle: 1 second
CT ratio (Number of turns):
1 to 1000
CTL-6-P-N: 800
CTL-12-S56-10L-N: 1000

Table of Contents

Other manuals for RKC INSTRUMENT RB100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKC INSTRUMENT RB100 and is the answer not in the manual?

RKC INSTRUMENT RB100 Specifications

General IconGeneral
BrandRKC INSTRUMENT
ModelRB100
CategoryTemperature Controller
LanguageEnglish

Related product manuals