Basic Instrument Functions
R&S
®
CMW500
87User Manual 1173.9463.02 ─ 06
Fig. 4-6: Setup - Remote
Most connection parameters are fixed and displayed for information only. The most
important parameters are described below.
Visa Resource (all protocol types)
The "Visa Resource" string depends on the protocol type and the assigned address
information. Each remote channel is identified by a different VISA resource string.
See also chapter 5.1, "Remote Control Operation", on page 117
Remote command:
SYSTem:COMMunicate:HISLip<i>:VRESource?
SYSTem:COMMunicate:SOCKet<i>:VRESource?
SYSTem:COMMunicate:USB:VRESource?
SYSTem:COMMunicate:VXI<i>:VRESource?
SYSTem:COMMunicate:GPIB<i>:VRESource?
Assigned Instrument (all protocol types)
Identifies the sub-instrument to be controlled. This parameter is fixed. If only one sub-
instrument is available, it is addressed as "Assigned Instrument 1".
If two sub-instruments are available, they are addressed as "Assigned Instrument 1" and
"Assigned Instrument 2". See also chapter 3.4, "Sub-Instruments", on page 76.
Via USB only sub-instrument 1 can be controlled in the present software version. Use
another protocol type to control sub-instrument 2.