Block H : AUTO Zero Set
H01 AUTO Zero Active
Range: NO
YES
Default: NO
For more information,
see Calibration item.
Reference: Nominal Belt Load D 01
Lets you select the automatic zero setting function
used to correct zero point on empty conveyor belt
after every belt circuit.
Prerequisites: v > vmin
No setting program
No parameter input
H02 Limit Value Diff.
Range: 0.00...10.00 %Q Default: 0.50 %Q
First prerequisite for automatic zero setting.
Defines the uniformity of mean belt load between two
consecutive belt circuits.
H03 Mean Limit Value
Range: 0.00...10.00 %Q Default: 1.00 %Q
Reference: Nominal Belt Load D 01
Second prerequisite for automatic zero setting.
Defines the maximum admissible belt load mean value
over 2 consecutive belt circuits.
H04 Zero Set Limit
Range: 0.00...100.00 %Q Default: 5.00 %Q
Reference: Nominal Belt Load D 01
Maximum admissible totals value of all manual and
automatic zero setting operations. Correction is not
acquired.
H05 Tare Corr. > MAX C5
Range: WARNING 1
ALARM
Default: WARNING 1
Automatic zero setting has exceeded Zero Set Limit
H04. Value is not acquired.
H06 Zero Set Time
Range: 1...999 h Default: 4 h
Lets you know that no manual or automatic zero
setting nor taring has been performed during the set
time.
Message C 7: Set Time To Zero
H07 Time Elapsed C7
Range: IGNORE
WARNING 1
WARNING 2
Default: IGNORE
Zeit : H 06
0837
INTECONT PLUS VBW
Parameterization
BV-H2214 GB 59