H08 Print Result
Range: NO
YES Default: NO
Print zero setting result after every belt circuit.
H09 ZDO Active
Range: NO
YES Default: NO
Activates Zero Drop Out function (see H 10).
H10 ZDO Limit
Range: 0.0...10.00 %Q Default: 1.00 %Q
Reference: Nominal Belt Load D 01
If ZDO function is active and belt load exceeds the
defined limit, feed rate measurement and totalization
are suppressed.
No symbol appears on display.
H11 Zero Set Diff.
Range: 0.00...100.00 %F Default: 5.0 %F
Reference: Nominal Belt Load D 01
Indicates the difference between two consecutive
manual or automatic zero setting operations.
Excess is reported by message C9.
H12 Tare Diff. > MAX C9
Range: WARNING 2
WARNING1
IGNORE
ALARM Default: WARNING 2
Manual or automatic zero setting has exceeded
difference H 11.
Correction is acquired save in case of ALARM
messages.
0837
VBW INTECONT PLUS
Parameterization
BV-H2214 GB 60
Technische
Redaktion PDE-RD
Transmission to third parties and reproduction of
this documentation are not permitted. Schenck
Process GmbH reserves all rights of ownership
and copyrights.
Result autom. zero setting
Total Tare [kg/m] 50,23 Tare [kg/m] 1,56
(D 03) (D 05)