7 Wavelength Calibration "FACTORY"
7.1 Outline
Automatic check function for wavelength calibration and wavelength accuracy
Wavelength calibration operation procedure
Use "FACTORY" - "
WAVE ADJ
" to display the measurement values (i.e. variance) of automatic
wavelength calibration and the wavelength after calibration.
With "CALIBRATION" - "WAVE CALIB", "CHECK GOOD" is displayed if the automatic
wavelength calibration and wavelength accuracy satisfy the criteria (±2.0 nm), and one of
"CHECK NO GOOD1/2/3" is displayed depending on the cause if these exceed the criteria.
This chapter explains "FACTORY" - "
WAVE ADJ
".
7.2 Preparations
One of water, methanol or acetonitrile (HPLC grade or
equivalent)
Low-pressure Hg (Mercury) lamp (RF-20A only)
Low-pressure Hg (Mercury) lamp holder (RF-20A only)