EasyManua.ls Logo

Shimadzu RF-20A - FACTORY Parameters; 10.1 FACTORY parameter list

Shimadzu RF-20A
91 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
49
10 FACTORY Parameters
10.1 FACTORY parameter list
As for the parameter to which F was attached, only FACTORY mode is displayed.
No.
Screen Display
Settable Range
Description
Remarks
1
ch1 EXxxx EMxxx
200 to 900 nm
CH1 EX/EM setting screen
2
ch2 EXxxx EMxxx
200 to 900 nm
CH2 EX/EM setting screen
3
ch3 EXxxx EMxxx
200 to 900 nm
F
CH3 EX/EM setting screen
4
ch4 EXxxx EMxxx
200 to 900 nm
F
CH4 EX/EM setting screen
5
MODE x
1-4
1: Single 2: Dual
Single-wavelength/Dual-wavelength selection
screen
Do not set 3 or 4.
6
LAMP x
Input 0 or 1
0: OFF 1: ON
Lamp lighting setting screen
7
RESPONSE x
Input 0 to 10
Response setting screen
8
SENS x
1: HI 2: MED 3:
LOW
F
Sensitivity setting screen
0: No filter, 1: 0.05, 2: 0.1, 3: 0.5, 4: 1.0, 5: 1.5,
6:3.0, 7:6.0, 8:8.0, 9:10.0, 10:2.0
9
SENS.
ENTER TO
START
F
Sensitivity calibration start screen
See 8.2
Compensating and
Verifying Sensitivity
at the Factory
(Reference)
10
-HV SENS.1 xxx
1-1000
F
Sensitivity 1 (HI) direct input screen
11
-HV SENS.2 xxx
1-1000
F
Sensitivity 2 (MID) direct input screen
12
-HV SENS.3 xxx
1-1000
Sensitivity 3 (LOW) direct input screen
13
GAIN x
1: 1 2: 4 3: 16
Gain setting
14
ANALOG1 MODE
x
Input 0, 1, 8, or
9
Analog 1 output mode setting
8: Cell temperature, 9: Room temperature
output (0C =0 mV to 100C =10 mV)
15
ANALOG2 MODE
x
Input 0, 1, 8, or 9
Analog 2 output mode setting
8: Cell temperature, 9: Room temperature
output (0C=0 mV to 100C =10 mV)
16
ANA1 REC
RANGE x
Input 0 to 9
Analog 1 range setting
17
ANA2 REC
RANGE x
Input 0 to 9
Analog 2 range setting
18
Z WAVE x
0: AZ, 1: F, 3: 3
to 5
F
0: Auto zero ON/1: Auto zero OFF (Other
settings are ignored.)
19
RATIO MODE x
0-2
F
0=S only output, 1=S/R both output (default),
2=R only output
S: Sample side, R: Reference side
20
WAVE ADJ
ENTER TO
START
F
Performs wavelength calibration. In the case of
the 20A, the Mercury lamp jig is required. After
calibration, the wavelength check is performed.
The wavelength calibration/check are performed
continuously at the
EX254 nm/EX507 nm/EX761 nm/
EM254 nm/EM507 nm/EM761 nm settings, and
finally the resulting wavelength variance is
displayed.
21
LOCAL x
0: Remote 1:
Local
Remote/local setting
22
LINK ADRS x
Input 1 to 12
Link address settings
23
SMPL EN
REF EN
12
186
Monitor: Sample energy
Monitor: Reference energy

Table of Contents

Related product manuals