52
output, 6: 10 mV output, 7: 100 mV output, 8:
500 mV output, 9: 1000 mV output, 51: Triangle
wave A (up to 1 mV in 1 minute, down to 0 V in
1 minute) (conditions: RATIO "0", response "0"),
52: Triangle wave B (up to 1 mV in 60 minutes,
down to 0 V in 60 minutes) (conditions: RATIO
"0", response "0"), 53: Triangle wave C (up to
1 mV in 720 minutes, down to 0 V in
720 minutes) (conditions: RATIO "0", response
"0")
can be set, never
enter parameters
other than the values
on the left. If a
parameter is entered
by mistake, enter "0"
and restart the
instrument.
48
TEMP CHECK
Enter to Check
R=30 S=33
OK
32.9-33.1
dT0.2
F
A check of a temperature control function is
possible, without using other external apparatus
(LabSolution, AD translation machine, etc.).
A result is automatic and is displayed.
Reference: 10.2.1
temperature test
mode [TEMP
CHECK] operation
procedure
49
WAVE CHECK
ENTER TO
START
F
Performs the wavelength check. The wavelength
check is performed continuously at the
EX254 nm/EX507 nm/EX761 nm/
EM254 nm/EM507 nm/EM761 nm settings, and
finally the resulting wavelength variance is
displayed.
Used for the
wavelength check