B:Adjustment of the emission-side grating position was added, and adjustment of the internal Mercury
lamp reflection mirror was changed.
C:Flow chart for Optical Adjustment was added., New PTFE block It is related with use of JIG and is
photograph change of the image of light.
D: 6.2 6. Clause Clerical Error Correction (on -> abut)15.6.4 (5) Clause Addition ([Z WAVE] Is
Changed into 0)
F: It is made in agreement with a Japanese amendment sign. 6.6 Correct EM PULSE value of the
table of [the relation between a reference:pulse and wavelength]. 6.11 (6) It is "etc. as before "... A
postscript is added. 6.12 (7)” Finally attach a cover etc. " is deleted.
G:Additional functional printing by V1.06 (temperature control test mode, automatic adjustment mode
of the reflective mirror for Hg lamps). P / N change of a cell gasket. "6.9 Adjustment of the
fluorescence mirror 1" is procedure correction a part.
H: P/N of "4.3.2 use parts list PCB, RF20-4.3.2 CPU" is corrected.
J: GAIN value of "15.6.3 Adjusting [Sens1],15.6.4 Adjusting [Sens3]" is changed.
K: Updated P/Ns in 13.2.1 Optical System Maintenance Parts List.
Updated P/Ns in 13.2.2 Flow Cell/Piping Parts Maintenance Parts List.
Added 16. Base ASSY, XS (including Peltier Unit) Replacement Procedure.
L: Parts number change of Photomultiplier of a maintenance part, and Lens,Cell EM.
M: Power Supply In connection with change, a part of 13.2.3 Electrical and Other Parts List was
changed.
N: New Mercury Lamp and a new Mercury Lamp holder are added to Chapter 6.3 List of Jigs.
P / N change of Mercury Lamp of Chapter 13.2.1 Optical System Maintenance Parts List.