10.2.1 Sensor diagnostic events
Sensor diagnostic events
ID/
Icons
Diagnostic Action Comments
32 SIL parameters not validated Run the 'Safety Validation' wizard to validate the
safety-critical parameters. The device can be
put into Safe Operation mode after validation
33 SIL parameters not validated Run the 'Safety Validation' wizard to validate the
safety-critical parameters. The device can be
put into Safe Operation mode after validation
34 SIL parameters not validated Run the 'Safety Validation' wizard to validate the
safety-critical parameters. The device can be
put into Safe Operation mode after validation
35 SIL parameters not validated Run the 'Safety Validation' wizard to validate the
safety-critical parameters. The device can be
put into Safe Operation mode after validation
36 Sensor supply volt. out of range Contact your local Siemens representative
37 Sensor supply volt. out of range Contact your local Siemens representative
38 Temperature measurement fault Turn off the power, wait 5 seconds and turn on
the power again. If the failure continues then
contact your local Siemens representative
39 Temperature measurement fault Turn off the power, wait 5 seconds and turn on
the power again. If the failure continues then
contact your local Siemens representative