Default settings
B
The following tables show the default settings as well as the range for various parameters.
The parameters can be accessed on three levels.
● Read Only
Allows no configuration. The user is only able to view the parameter values. No PIN code
required.
● User
Allows configuration and service of all parameters except calibration parameters. Default
PIN code is 2457.
● Expert
Allows configuration and service of all parameters including flow and density calibration
parameters. Default PIN code is 2834.
B.1 Sensor
Table B-1 Sensor
HMI menu ID Parameter Description Default value
[unit]
Value range Access
level
2.1.1 Flow Direction Select pos/neg flow. Default pos
flow is indicated by arrow on sen‐
sor.
Positive
● Negative
● Positive
User
2.1.2 Process Noise Damping Select process noise damping
level.
* 1: Low … 5: High
Duplex pump (3)
● Centrifugal
pump (1:low)
● Triplex pump
(2)
● Duplex pump
(3)
● Simplex pump
(4)
● Cam pump (5)
high
User
2.1.3 Zero point adjustment
2.1.3.1 Select Zero Point Adj. Selected zero-point adjustment
method. Automatic zero point
adjustment is recommended.
Auto
● Auto
● Manual
2.1.3.2 Zero Point Adjustment Wizard
2.1.3.3 Duration Define zero-point adjustment du‐
ration.
30
[s]
1 to 999 User
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA 271