B.5 Inputs and outputs
Table B-10 Channel 1 – output with HART
HMI menu ID Parameter Description Default value
[unit]
Value range Access
level
2.4.1.1 Loop current mode Select single point or multi-drop
HART communication.
Loop current
mode, 4..20 mA
● Off
● Loop current
mode, 4..20
mA
● Multi drop
mode, 4 mA
User
2.4.1.2 Active operation Shows the application possibili‐
ties of your hardware. The kind
of operation depends on the wir‐
ing.
Active operation
not possible
● Active
operation not
possible
● Active
operation
possible
User
2.4.1.3 PV selection Select process value for current
output.
Mass flow For none function‐
al safety applica‐
tions (SIL):
● Mass flow
● Volume flow
● Standard
volume flow
● Density
● Reference
density
● Medium
temperature
● Fraction A
● Fraction B
● Fraction A %
● Fraction B %
● Control valve
For functional
safety applica‐
tions (SIL):
● Mass flow
● Volume flow
● Density
User
Default settings
B.5 Inputs and outputs
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA 291