6. Mount power supply protection cover and fasten protection cover screw.
7. Tighten cable gland.
5.2.4.8 Step 4b: Connecting the inputs and outputs (channels 2 to 4)
1. Remove cap and ferrule from cable gland and slide onto cable.
2. Push cable through open gland and cable path.
3. Restore ferrule and tighten cap to lightly hold cable in place.
4. Fold signal cable screen back over outer sheath and ground beneath cable clamp.
5. Connect wires to terminals using wiring tool.
11
12
12
13
Active current output Passive current output
IO[3] (common) IO[3]- (passive)
IO[3]+ (active) IO[3] (common)
Functional Earth Functional Earth
Termination example for channel 3
6. Tighten cable gland.
Note
Active or passive current output is preselected at ordering.
Connecting
5.2 Connecting MASS 2100/FC300
MASS 2100 & FC300 (FCT030)
78 Operating Instructions, 06/2017, A5E39789040-AA