ID/
Icons
Diagnostic Action Comment
195 CH2 Loop current in lower saturation. Check process
conditions or align limit to normal operation. Ad‐
just channel 2 parameter 'Lower range value'
If Operation mode is config‐
ured to Current output
196 CH2 Loop current in upper saturation. Check proc‐
ess conditions or align limit to normal operation.
Adjust channel 2 parameter 'Upper range value'
If Operation mode is config‐
ured to Current output
197 CH2 Cable break. Check channel 2 current output
cable connection.
If Operation mode is config‐
ured to Current output
198 CH2 Frequency too low. Check process conditions
or align limit to normal operation. Adjust chan‐
nel 2 parameter 'Lower range value'
If Operation mode is config‐
ured to Frequency output
199 CH2 Frequency too high. Check process conditions
or align limit to normal operation. Adjust chan‐
nel 2 parameter 'Upper range value'
If Operation mode is config‐
ured to Frequency output
200 CH2 Pulse overflow. Pulse output insufficient pulse
separation. Reduce pulses per amount, or re‐
duce pulse width, or increase amount
If Operation mode is config‐
ured to Pulse output
201 CH3 Loop current in lower saturation. Check process
conditions or align limit to normal operation. Ad‐
just channel 3 parameter 'Lower range value'
If Operation mode is config‐
ured to Current output
202 CH3 Loop current in upper saturation. Check proc‐
ess conditions or align limit to normal operation.
Adjust channel 3 parameter 'Upper range value'
If Operation mode is config‐
ured to Current output
203 CH3 Cable break. Check channel 3 current output
cable connection
If Operation mode is config‐
ured to Current output
Diagnosing and troubleshooting
10.2 Fault codes and corrective actions
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA 205