For US:
FCT030 field mount transmitter (can be installed in Zone 1 for gas and Zone 21 for dust) and
Class I +II+III Group A, B, C, D, E, F:
Certificate: cCSAus 2508628
Class I, II, III Division 1 Gp A, B, C, D, E, F, G
Class I Zone 1: AEx db eb ia [ia Ga] IIC T6 Gb
Class II Zone 21: AEx tb [ia Da] IIIC T85°C Db
Installation variations
Note
Requirements for safe installation
Standard remote installation with FCT030 because the connection is certified Intrinsically
Safe. however flameproof seals and conduit (for IS cable) can be used.
Requirement for IS circuit is that the maximum input voltage Vi to DSL is 20 VDC, Ii is
maximum 484 mA, Pi < 2.3 W
FC300
ATEX Certificate : DEMKO 05 ATEX 138072X.
Sitrans FC300 can be used in zone 0.
Ex ia IIC T6…T3 Ga
IECEx Certificate: IECEx ULD 17.0030
Sitrans FC300 can be used in zone 0.
Ex ia IIC T6…T3 Ga
Canada Certificate. UL E232147
Sitrans FC300 can be used in zone 0 and Class I Division 1.
Class I Div. 1 Grp. A, B, C, D.
Ex ia IIC T6…T3 IIC Ga
US Certifikate. UL E232147
Sitrans FC300 can be used in zone 0 and Class I Division 1.
Class I Div. 1 Grp. A, B, C, D
AEx ia IIC T6…T3 IIC Ga
Safety notes
2.4 Use in hazardous areas
MASS 2100 & FC300 (FCT030)
22 Operating Instructions, 06/2017, A5E39789040-AA