Zero Point Adjustment
Sensor type Sensor size Default value
Kg/s
Standard Deviation Limit
FC400 DN 15 0,0004
DN 25 0,004
DN 50 0,015
DN 80 0,036
DN 100 0,0832
DN 150 0,1376
MASS2100 1 0
1.5 0,000002
2.1 0,000002
3 0,00002
FC300 DN4 HP 2 0
FC300 DN4 C22 3 0,00002
FC300 DN4 316 3.5 0,00002
MASS2100 6 0,0001
15 0,0004
Zero Offset Limit
FC400 DN 15 0,0103
DN 25 0,0319
DN 50 0,1444
DN 80 0,378
DN 100 0,783
DN 150 1,269
MASS2100 1 0,0000272
1.5 0,0000542
2.1 0,000117
3 0,000250
FC300 DN4 HP 2 0,0000972
FC300 DN4 C22 3 0,000264
FC300 DN4 316 3.5 0,000389
MASS2100 6 0,00139
15 0,0106
Technical reference
A.1 Sensor dimension dependent default settings
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA 267