HMI menu ID Parameter Description Default value
[unit]
Value range Access
level
4.2.1 Polling address (SW) Enter the device HART address.
HW address device address
overwrites the SW device ad‐
dress.
0 0 to 63 User
4.2.2 Polling address (HW) Hardware device address. The
HW device address is config‐
ured by DIP switch in the device.
0 0 to 16 Read
only
4.2.3 TAG Enter a unique Short TAG name
for the device / measurement
point.
Up to 8 characters User
4.2.4 HART device Type Device Identification: Uniquely
identifies the Field Device when
combined with the Manufacturer
Identification and Device Type.
34 Read
oSLy
4.2.5 HART revision Device Revision Level defined in
HART protocol.
0 Read
oSLy
4.2.6 Response preambles Response preambles 5 5 to 20
4.2.7 Dynamic variable mapping
4.2.7.1 SV process value Select the Secondary HART
process value.
Volume flow
Mass flow
Volume flow
Standard
volume flow
Density
Standard
density
Medium
temperature
Fraction A
Fraction B
Fraction A %
Fraction B %
Totalizer 1
Totalizer 2
Totalizer 3
User
Default settings
B.13 Communication
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA 467