Factory configu‐
ration
Software configuration Channel 2 Channel 3,
Channel 4
Relay output Alarm Class
Individual alarms
One on/off valve
Two on/off valves
X
8
H[W
+
-
ChXC
ChX+
ChX-
Normally open
Relay output Alarm Class
Individual alarms
One on/off valve
Two on/off valves
X
8
H[W
+
-
ChXC
ChX+
ChX-
Normally closed
Load
Ensure the total load connected to the output is sufficient to obtain the voltage level required
by the module.
Load [Ω] Voltage (active
Ex) [V]
Voltage (active non
Ex) [V]
24 V DC Voltage supply (passive) [V] Ex and
Non Ex
Measurered
100 3 3 17.7
200 5.9 5.9 19.6
500 11.4 14.9 21.0
1000 14.8 19.2 21.6
2000 17.4 20.1 21.9
5000 19.4 20.4 22.2
10000 20.3 21 22.3
20000 20 20.4 22.4
50000 20.4 20.6 22.5
100000 20.6 20.7 22.6
Note
Load
Signal output: < 500 Ω at 14 to 24 VDC (active), 14 to 30 VDC (passive)
Relay output: 30 VAC/VDC, 100 mA
Passive signal input: 15 to 30 VDC, 2 to 15 mA
Connecting
5.2 Connecting MASS 2100/FC300
MASS 2100 & FC300 (FCT030)
80 Operating Instructions, 06/2017, A5E39789040-AA