Functions
8.3 Calibration
FIDAMAT 6
86 Operating Instructions, 01/2019, A5E00222135-04
solenoid valves. The same also applies to the "AUTOCAL" signaling contact. This is closed
for approximately 1 s when the command is executed.
Example:
You want to program the following sequence:
1. Zero gas calibration, after 15 minutes of purging with zero gas
2. Calibration with calibration gas 1, after 10 minutes of purging with calibration gas
3. Purging with sample gas: 8 minutes
4. Sample gas intermediate operation: 30 minutes
5. Calibration with calibration gas 2, after 8 minutes of purging
6. Calibration with calibration gas 3, after 8 minutes of purging
7. Calibration with calibration gas 4, after 10 minutes of purging
8. Purging with sample gas: 8 minutes
9. Short-term signaling contact in order to be able to start AUTOCAL on another device or
channel.
The specified AUTOCAL sequence is shown in the following function screens.
Figure 8-11 Example AUTOCAL sequence
Figure 8-12 Example AUTOCAL sequence